OPERATION
57
CV Mode
In Constant Voltage Mode, the unit will maintain a
constant voltage. In CV mode you set the constant
voltage level. For more details on CV mode, see the
appendix on page 226.
If you change the mode or the range when the load is
already on, the load will be turned off automatically.
1. Make sure the load is off.
2. Press
.
3. Select CV mode with the Mode[F1] soft-key.
4. Select the current range with the I Range[F2]
soft-key.
5. Select the voltage range with the V Range[F3]
soft-key.
6. Set the voltage level parameters using the scroll
wheel and number pad.
Set CV A Value and/or CV B Value.
The maximum and minimum voltage levels
depend on the selected voltage range.
7. Set the remaining basic configuration settings
such as the response settings. See page 65 for
details.