PEL-3000(H) User Manual
96
5. Set the Delay Time.
The delay time setting will delay activating the
Go-NoGo testing by a specified amount of time.
The delay setting can compensate for startup
oscillation and other instabilities during
startup.
Delay Time
0.0-1.0 seconds (0.1s resolution )
When the Main settings are saved or recalled, the Go-
NoGo settings are also saved/recalled. See the
Save/Recall chapter for details, page 98.
Running a Go-NoGo Test
Description
Go-NoGo test results are displayed in the
measurement panel.
GO indicates pass (good).
NG indicates fail (no good).
Operation
1. Press
Main
> Configure[F5] > Go-NoGo[F3].
2. Set SPEC Test to ON.
When SPEC Test is ON, SPEC will appear in the
Operation Status Panel. This means the unit is
ready for Go-NoGo testing.
3. Turn the load on.
The test starts from the time the load was
turned on + the Delay Time.
Display:
GO
0.000
V
0.0000
A
0.00
W
Level2 0.00
Level1 0.00
LOADRS232
01/Oct/2018
W
W
SPEC test = ON
SPEC
GO
Test result