EasyManua.ls Logo

GW Instek PEL-3000 Series - Trigger out Width; Measure Average

GW Instek PEL-3000 Series
257 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PEL-3000(H) User Manual
94
Operation
1. Press
Shift
+
Utility
Help
> Other[F5].
2. Set the Trig In Delay setting.
Range:
0.0 - 5000µs
Default: 0µs
Trigger Out Width
Description
The Trigger Out Width setting sets the trigger
output signal’s pulse width.
Operation
1. Press
Shift
+
Utility
Help
> Other[F5].
2. Set the Trig Out Width setting.
Range:
2.5 - 5000.0µs
Default: 10µs
Measure Average
Description
The Measure Average setting is used to set the
speed of the measurement display. The setting has
three modes. They are slow, normal and fast
The default mode for Measure Average setting is
slow.
Operation
1. Press
Shift
+
Utility
Help
> Other[F5].
2. Set the Measure Average setting.
Slow
Normal
Fast
Default
Average 64 times; Display spend time:1280ms
Average 16 times; Display spend time:320ms
Average 4 times; Display spend time:320ms
Slow mode