OPERATION
75
Von Delay
Von Delay is the amount of time the unit will wait
before turning the load on after the Von Voltage
threshold has been latched. This will prevent
overshoot current from affecting the Von Voltage
threshold.
> Configure[F5] > Other[F2] and
set the Von Delay time.
Von Delay: OFF, 2.0-60ms(CC, CV, CP
mode)
OFF, 5.0-60ms (2.0-60ms)(CR
mode)
CR mode can have the delay time set separately from
the other modes (called
Von Delay –CR
when in CR
mode).
Timer Functions
Count Time
When Count Time is set to on, it will count the
elapsed time from when the load was turned on to
when it was turned off.
This function is applicable to manual and
automatic shutdown (such as from protection
functions such as UVP etc.)
The elapsed time will be shown in the display
Measurement area.
> Configure[F5] > Other[F2] and
turn the Count Time on or off.