OPERATION
87
OVP
If the OVP is tripped, the PEL-3000(H) will turn off
the load.
The OVP levels can be set from 0V to 10% higher
than the rating voltage.
> Configure[F5] > Protection[F1]
and set the OVP Level.
OVP Level: OFF, 0-Rating voltage + 10%
To turn OVP off, set the OVP voltage greater than the
current rating voltage + 10%.
The OVP indicator and a message will only
appear on the screen when the input voltage is
below the UVP level. The Enter key must be
pressed to clear the alarm message.
To clear the OVP indicator, remove the cause of
the over voltage - i.e., reduce the input voltage.