PEL-3000(H) User Manual
72
Advanced Configuration Settings
Use the advanced configuration settings to configure settings other
than those described in the basic configuration chapter.
Soft Start
Description
The soft start setting is used to limit the amount of
input current at start-up or from when the Von
Voltage threshold is tripped.
The soft start setting only applies to CC, CR
and CP mode.
Input I
Time
Soft Start = ON
Input I
Time
Soft Start= OFF
Rise voltage
Rise current
Rise current
Rise voltage
Operation
1. Press
Main
> Configure[F5] > Other[F2] and
set the Soft Start time.
Range:
OFF, 1-200ms