PEL-3000(H) User Manual
244
Reverse voltage protection(RVP)
By diode. Turns off the load when an alarm occurs.
Rating overcurrent protection (ROCP)
An ROCP message will be produced when the input current range
is greater than 110% of the rated operating current range (I
range).
Rating overpower protection (ROPP)
An ROPP message will be produced when the input power range
is greater than 110% of the rated operating power range.
Front panel input rating overcurrent protection (F.ROCP)
An F.ROCP message will be produced when the front panel input
current range is greater than 77A (Typical).
Sequence
0.05 ms (0.05 ms – 1 min)/100 ms (1 min – 1 h)/1 s
(1 h – 10 h)/10 s (10 h – 100 h)/1 min (100 h – 999
h 59 min)
1us(25us -60ms) /10us(60.01ms -600ms)
Other
Measures the time from load on to load off. On/Off selectable.
Measures from 1 s up to 999 h 59 min 59 s
Automatically turns off the load after a specified time elapses.
Can be set in the range of 1 s to 999 h 59 min 59 s or off
IEEE std. 488.1-1978 (partial support)
SH1, AH1, T6, L4, SR1, DC1, DT1.
Supports the SCPI and IEEE std. 488.2-1992 command set
Sets panel functions except the power switch and reads measured
values