EXTERNAL CONTROL
191
MASTER
+
-
+S
-S
SLAVE 1
+
-
+S
-S
+
-
+S
-S
SLAVE 2
+
-
+S
-S
DUT
-
+
J2
J1
J2
J1
J2
J1
J2
J1
SLAVE 3
Load
wiring
Frame link cable
(GTL-255)
Sense
wiring
Ferrite cores
No ferrite core to
last J1 connection
Cautions
Only the rear terminals can be used for parallel
connections.
Make sure all connections are correct before turning
on the load. Incorrect connections could damage the
units.
Only units of the same type and rating can be used in
parallel (except for when the PEL-3211(H) booster
pack is used with the PEL-3111(H)).
Ensure that wiring of sufficient gauge is used when
using parallel connections.
If using remote sense, only connect the master to the
voltage sense terminals.