PEL-3000(H) User Manual
66
Level
Time
Level 2
Level 1
Timer1 Timer2 Timer1
Dynamic Mode
Dynamic mode is not available for CV mode.
Operation
1. Make sure the load is off.
2. Press
Main
.
3. Select Dynamic or Static mode with the
Function[F4] soft-key.
A different switching function can be set for
CC, CR and CP mode.
4. For dynamic mode, set the Timer1 and Timer2
parameters using the scroll wheel and number
pad.
Timer1 sets the Level1 on-time.
Timer2 sets the Level2 on-time.
Take the slew rate settings into consideration
when setting the timers.
The frequency of the dynamic switching is
output via the TRIG OUT BNC.
5. In the static mode, select either value A or value
B as the active value for execution followed
by pressing the
Shift
+
Preset
.
The “active” value will be shown in the
Operation Status Panel.
The load can be “on” when switching between
A Value and B Value.