9705 - System Setup 27 of 82
KSS - March 2006 - Edition 1
HIAC Operator Manual
o
e:
Use the << and >> keys to move the cursor through the field; the DELETE and BKSPC keys
will not work. Do not enter separators (/); they will automatically appear.
4) When the date is correct, press Enter. The keypad entry screen closes and the System
Settings screen displays.
5) Press Return to save changes and exit.
3.5 Setting Particle Sizes
From the Main Screen, select the Setup icon > Sizes icon to invoke the Size Settings screen,
shown in Figure 3-6.
These functions are accessed through the icons on this tab:
• Adding sizes
• Editing sizes
• Deleting sizes
The minimum size that may be entered is 1.3; the maximum size is 200.0.
3.5.1 Adding a Size
o
e:
The 9705 will not extrapolate beyond the calibration curves. Setting sizes outside the
calibration curves will result in an error.
1) From the Main Screen, select the Setup icon > Sizes icon to invoke the Size Settings
screen, shown in Figure 3-6.
2) Touch the Add Size icon from the Sizes tab to invoke the Particle Size Entry screen.
3) Use the touch screen to enter a new size.
Fig 3-6 : Sizes Tab