4) Connect the tube to the sampling probe. Tighten it completely by hand.
CAUTION:
The tube is to be inserted and held flush with the bottom of the port while the ferrules are
pushed in and the nut is tightened.
5) Remove the tube from the sampling probe and inspect the ferrule crimp. It should look
the same as the ferrule cut off in step 1. The tube end should be flush with the ferrule
end and free of nicks and burrs.
Fig 3-34 : Connect the Tube to the Probe
Fig 3-35 : Tube End Flush with Ferrule End