32 of 82 System Setup - 9705
KSS - March 2006 - Edition 1
Operator Manual HIAC
3) Press the Return icon to save changes and return to the main screen.
3.9 Changing the Needles
The 9705 comes with two vent needles to allow sampling from a variety of bottle sizes.
The shorter vent needle is used to sample 5-30 mL bottles.
The longer vent needle is used to sample 50-125 mL bottles.
The vent needle must be removed to use the Large Volume Adapter (LVA); this process is
described in “Connecting the Large Volume Adapter” on page 35.
N
o
t
e:
The short sampling needle does not need to be removed at any time. Do not attempt to
remove the shortest needle.
To change needles:
1) Tilt the bottle dock forward.
2) Remove the bottle guide.
Fig 3-10 : Tilt Bottle Dock