Calibration and Configuration 115
Haemonetics
®
ACP-215
®
Service Manual P/N: SM-ACP215-01-EN(AB)
Protocol configuration parameters are setup with multiple levels of interface
and are dependent on market use. The following tables will show each
configurable parameter, its default value, and available user interface.
Note: Once the market has been selected the default market values are set into
the memory of the machine as displayed in the market tables below.
Note: These parameters are market dependent and are only changed if the
“Machine Market” configuration is changed.
Table 7
Config Parameter Default Value Selections
Machine Market International International, North
America
Table 8, Configuration Parameter Default Settings
International Market North America Market
Config Parameter Default Value Config Parameter Default Value
*Disposable Checklist Minimum Set *Disposable Checklist Full Set
*Modifiable Parameters Full Set *Modifiable Parameters Minimum Set
Available Protocols Frozen Blood Available Protocols Frozen Blood
Cell Wash Additive Option Enabled Cell Wash Additive Option Disabled
Bowl Size Option Small Bowl
Only (275ml)
Bowl Size Option Small Bowl
Only (275ml)
*Sterile Check in Cell Wash Enabled *Sterile Check in Cell Wash Disabled