114 Calibration and Configuration
P/N: SM-ACP215-01-EN(AB) Haemonetics
®
ACP-215
®
Service Manual
19.Press MODIFY PROGRAM to move the pointer on the screen to the De-
Glycerolization protocol option parameter.
20.Press START. The screen displays:
Figure 34, Bowl size option in deglycerolization
21.Press MODIFY PROGRAM, if necessary, to move the pointer on the
screen to the Bowl size option parameter.
22.Press START. The screen displays:
23.Press MODIFY PROGRAM to change the of the Bowl Size setting within
the deglycerolization protocol if necessary, then press START to
confirm.
24.Press STOP five times to return to the Main Configuration screen.
GLYCEROLIZATION CONFIGURATION
International
Cell Wash
Glycerolization
-> De-Glycerolization
DEGLYCEROLIZATION CONFIGURATION
International
-> Bowl size option
BOWL SIZE OPTION PARAMETER
-> Small Bowl Only (275ml)
Large Bowl Only (325ml)
Both
Status = Small Bowl Only
CONFIGURATION
-> PROTOCOLS CONFIGURATION
LANGUAGE
MACHINE