Principles of Operation 45
Haemonetics
®
ACP-215
®
Service Manual P/N: SM-ACP215-01-EN(AB)
If the DPM and SPM are not exposed to ambient pressure, the transducer
signals may affect tests later in the sequence. So before testing the A/D
converter, the DPM and SPM interlock switches are polled to check for a
loaded disposable filter. If an interlock signal indicates a loaded disposable
filter, one of the following messages will be displayed and a warning beep will
sound once per second:
If the filter is not removed within fifty seconds, the interlock will be presumed
defective and the test will fail.
PLEASE CLAMP THE DPM LINE AND
DISCONNECT THE FILTER
PLEASE CLAMP THE SPM LINE AND
DISCONNECT THE FILTER
DPM FILTER INTERLOCK FAILURE
CODE = 507
PLEASE POWER OFF.
DPM FILTER INTERLOCK FAILURE
CODE = 507
PLEASE POWER OFF.