46 Principles of Operation
P/N: SM-ACP215-01-EN(AB) Haemonetics
®
ACP-215
®
Service Manual
If the A/D converter fails one of its tests, one of the following messages will be
displayed:
SPM FILTER INTERLOCK FAILURE
CODE = 502
PLEASE POWER OFF.
DPM FILTER INTERLOCK FAILURE
SPM FILTER INTERLOCK FAILURE
CODE = 504
PLEASE POWER OFF.
DPM FAILED 20% ANALOG TEST
CODE = 10
PLEASE POWER OFF.
DPM FAILED 50% ANALOG TEST
CODE = 10
PLEASE POWER OFF.
DPM FAILED 80% ANALOG TEST
CODE = 10
PLEASE POWER OFF.