80 Preventive Maintenance and Installation
P/N: SM-ACP215-01-EN(AB) Haemonetics
®
ACP-215
®
Service Manual
15.Press NO to stop the centrifuge test and continue to the next step. The
screen displays:
16.Press YES to continue. The screen displays:
17.Press YES to continue.
18.Repeat for the yellow, orange, red, clear, green, and line sensor tubing
screens. The screen displays:
19.Press YES to continue.
20.Repeat for the wash solution, additive solution, and solution air detector
screens.
DE-GLYCEROLIZATION PROTOCOL
Start Disposable Check List
Press YES To Continue
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
Check, Press YES To Continue
BLUE Striped Tubing in BLUE
Valve.
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
Check, Press YES To Continue
BLUE Striped tubing connected to
Hypertonic Solution bag.
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240