Preventive Maintenance and Installation 81
Haemonetics
®
ACP-215
®
Service Manual P/N: SM-ACP215-01-EN(AB)
21.Ensure that the water-filled tubing is laced in the solution air detector and
the light is on.
22.Press YES to continue. The screen displays:
23.Press YES to continue. The screen displays:
24.Press YES to continue. The screen displays:
25.Ensure that the water-filled tubing is laced in the blood air detector and
the light is on.
Check, Press YES To Continue
Solution Pump Loaded Correctly.
Filter Downstream Pump (CCW).
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
Check, Press YES To Continue
Blood Pump Loaded Correctly. Filter
Upstream Pump (CCW).
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240
Check, Press YES To Continue
Blood tubing Laced in
Blood Air Detector.
PUMP TIME HYPE NACL ADDI
REAL: 0 0 0 0 0
TARG: 0 56 50 1470 240