EasyManua.ls Logo

Haemonetics ACP 215 - Page 94

Haemonetics ACP 215
186 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
94 Preventive Maintenance and Installation
P/N: SM-ACP215-01-EN(AB) Haemonetics
®
ACP-215
®
Service Manual
6. Press PUMP UP ARROW and the valve closes.
Figure 29, Hidden keys on membrane panel
Test valve 2 (white)
1. Press PUMP START/STOP and valve 2 (white) opens.
2. Install a length 0.170” tubing through the open pinch valve 2 (white) and
attach one end to a sphygmomanometer or digital pressure meter.
3. Verify that valve 2 opens and its LED is lit.
4. Press PUMP START/STOP and the valve closes.
5. Verify that valve 2 closes and its LED is not lit.
6. Pressurize the sphygmomanometer or digital pressure meter to
200 mmHg and verify that there is not a leak as evidenced by a decrease
in pressure. Press PUMP START/STOP and remove the tubing.
7. Press PUMP START/STOP and the valve closes.
Test valve 3 (green)
1. Press PUMP DOWN ARROW and valve 3 (green) opens.
2. Install a length 0.170” tubing through the open pinch valve 3 (green) and
attach one end to a sphygmomanometer or digital pressure meter.
3. Verify that valve 3 opens and its LED is lit.
4. Press PUMP DOWN ARROW and the valve closes.
5. Verify that valve 3 closes and its LED is not lit.

Table of Contents

Related product manuals