Iss. a – Translation of the original version
Doc. Ref. : SOUND2.940.001
Valeur du spectre de puissance dans la zone de
mesure
Value of the power spectrum in the
measurement zone
1. Check the default values in the “Calibration” menu.
The 1
st
calibration point is carried out empty, in order to measure the sound when the mill is empty.
The “Power Max” must be other than 0 and vary moderately around a fixed value.
2. Press “Start” to set the 1
st
calibration point. The controller goes back to the main screen and the text
“Calibration in progress” flashes at the top of the display.
3. Wait for a few seconds. When calibration is complete, the text “Ready” is displayed and the Filling (or
Noise) level is indicated on the screen
Figure 27: “Calibration in progress” menu