FAQs: Frequently asked questions (MP85A and PME Assistant)
138 A0293_06_E00_03 HBM: public MP85A
RemedyCauseMessage
Initial calibration
error [InCalErr]
The factory calibra
tion of the MP85A is
faulty
This may be a one-off error; please reboot
(switch off and then on again after approx.
30 seconds). If the error reoccurs, please
contact HBM Service or HBM Support (see
PME Assistant online help).
Memory
allocation error
The internal memory
(RAM) no longer has
any space for
process data
This error occurs if the data is not being
copied fast enough or at all from the RAM
to the storage destination.
If you also see a status message saying
that the MMC/SD card is full, replace it.
If you also see a message saying that the
internal memory is nearly full, check
whether the data can be output fast
enough via the storage destination. If
necessary, reduce the scope of the output
data, or select a different storage
destination.
MMC/SD card
faulty [!]
The MMC/SD card
was not recognized,
or was recognized
incorrectly
If you have just inserted a new card, check
that it is the correct type (standard MMC)
and has the right format (FAT16).