123
Diluted sample
Please Fill
and
Insert a Cuvette
STAT Test
2
3
FIGURE 5-3
The Dilution function is only applicable for
HemoCue Glucose 201 DM Analyzer.
The function is not available in all markets.
Note that the Dilution button (2) and Help
button (3) will only be displayed if the
dilution option was chosen during the
configuration of the Analyzer settings (see
section 3 Settings).
For the Dilution function the sample must
be diluted with NaCl (Saline), one part blood
and one part NaCl.
f) If the sample is diluted, press the Dilution
button (2). FIGURE 5-4 will be displayed.
2 Dilution button
3 Help button
Dilution
NoNo YesYes
Sample Diluted
with NaCl 1+1?
FIGURE 5-4
g) Confirm with Yes or No. FIGURE 5-3 will
then be displayed again.
h) Obtain a blood sample according to the
procedure described in relevant
Instructions for Use.
i) Place the filled Cuvette in the Cuvette
holder and gently insert it into the
measuring position.
j) When the measurement is ready, FIGURE
5-5 will be displayed.

Table of Contents

Other manuals for HemoCue 201 DM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HemoCue 201 DM and is the answer not in the manual?

HemoCue 201 DM Specifications

General IconGeneral
BrandHemoCue
Model201 DM
CategoryDocking Station
LanguageEnglish

Related product manuals