44
Unit:mg/dL
Warning Min (0) 30
Warning Max (500) 60
1
5
2
87
4 6
9
3
0
11
55
22
8877
44 66
99
33
00
10
11
12
FIGURE 3-11
The warning limits define the range of
results for QC testing that must be obtained
to prevent a warning from being displayed.
j) Enter min. value(10) and max. value(11)
for the warning interval.
k) Press the Confirm button(12).
A similar image for the lockout limits will be
displayed. The lockout limits define the
range for QC testing that must be obtained
to prevent the Analyzer from performing a
lockout.
l) In the image for the lockout limits, enter
min. value(10) and max. value(11) for the
lockout interval.
m) Press the Confirm button(12).
10 Min. value
11 Max. value
12 Confirm button