EasyManua.ls Logo

HemoCue 201 DM - Page 78

Default Icon
203 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
73
Full Memory Options
LockoutLockout
OverwriteOverwrite
Patient
Memory Full
2
3
1
FIGURE 3-53
Up to 4000 Patient/STAT tests can be stored
in the Analyzer. When the Memory button is
pressed in the General Settings menu (see
FIGURE 3-46), the options for when the
patient memory is full are displayed.
The options available are displayed as
buttons(1). The option currently in use is
indicated by the selected Radio button(2).
Lockout: The Analyzer is locked out when
the patient memory is full. No patient tests
can be performed until the stored results
have been deleted from the Patient Memory
function.
Overwrite: When the patient memory is full,
the Analyzer stores the new patient test
result by overwriting the oldest result.
a) To change the setting, press the button
for the desired option(1).
b) Press the Confirm button(3).
FIGURE 3-54 will be displayed.
1 Full Memory options
2 Selected option
3 Confirm button

Table of Contents

Other manuals for HemoCue 201 DM

Related product manuals