88
QC Test settings for distinct concentration
levels
An Analyzer can be set to perform a lockout
if the required QC test is not performed
within a pre-defined time or after a pre-
defined number of measurements. The
Analyzer will be unlocked when an approved
QC test at the required concentration level
has been performed. The lockout
configuration must be set for each
concentration level of Liquid Control.
Not UsedNot Used
TimeTime
MeasurementsMeasurements
QC Level 1
Lockout
3
1
2
FIGURE 3-68
When the Level 1 button is pressed in the
QC Test Settings menu (see FIGURE 3-67),
the settings for QC lockout at this level are
displayed.
The options available are displayed as
buttons. The option currently in use is
indicated by the selected radio button.
Not Used: No QC lockout will be performed.
Time: The QC lockout will be activated if the
required QC test has not been performed
within the defined time interval.
Measurements: The QC lockout will occur if
the required QC test has not been
performed within the predefined number of
measurements.
1 Options for QC lockout criteria
2 Selected option
3 Confirm button

Table of Contents

Other manuals for HemoCue 201 DM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HemoCue 201 DM and is the answer not in the manual?

HemoCue 201 DM Specifications

General IconGeneral
BrandHemoCue
Model201 DM
CategoryDocking Station
LanguageEnglish

Related product manuals