Specifications
Chapter 1: Introduction
Basic Care™ Bed Service Manual (man336) Page 1 - 7
1
Radiated RF
IEC 61000-4-3
3 Vrms
80 MHz to 2.5 GHz
3 Vrms
80 MHz to 2.5 GHz
Portable and mobile RF
communications equip-
ment should not be used
at close distances to the
Basic Care™ Bed. (See
Note 2.)
Electrical Fast
Tran-
sient/Burst
IEC 61000-4-4
± 2 kV on Power
Supply Lines
± 1 kV on Input/
Output Lines
± 2 kV on Power
Supply Lines
± 1 kV on Input/
Output Lines
Mains power quality
should be that of a typi-
cal commercial or hospi-
tal environment.
Surge
IEC 61000-4-5
± 1 kV Differential
Mode
± 2 kV Common
Mode
± 1 kV Differential
Mode
± 2 kV Common
Mode
Mains power quality
should be that of a typi-
cal commercial or hospi-
tal environment.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
10 Vrms from 80
MHz to 2.5 GHz
3 Vrms
150 kHz to 80 MHz
10 Vrms from 80
MHz to 2.5 GHz
Portable and mobile RF
communications equip-
ment (cell phones)
should not be used at
close distances to the
Basic Care™ Bed. (See
Note 2.)
Power Fre-
quency Mag-
netic Fields
IEC 61000-4-8
3 A/m 3 A/m Power frequency mag-
netic fields should be at
levels characteristic of a
typical location in a typi-
cal commercial or hospi-
tal environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment—Guidance