2.2 The Head Up Function Does Not Operate
Chapter 2: Troubleshooting Procedures
Basic Care™ Bed Service Manual (man336) Page 2 - 7
2
8. Replace the controls on the siderail with the faulty function. The head up
function operates from this siderail.
Yes No
↓→Replace the siderail with the faulty function. If this corrects the
problem, go to “Final Actions” on page 2-4. Otherwise, call
Hill-Rom Technical Support at 1-812-934-1796.
9. Go to “Final Actions” on page 2-4.