2.6 The Knee Down Function Does Not Operate
Chapter 2: Troubleshooting Procedures
Page 2 - 14 Basic Care™ Bed Service Manual (man336)
8. Replace the controls on the siderail having the knee down function that
does not operate. The knee down function operates from this siderail.
Yes No
↓→Replace the siderail having the knee down function that does not
operate. If this corrects the problem, go to “Final Actions” on
page 2-4. Otherwise, call Hill-Rom Technical Support at
1-812-934-1796.
9. Go to “Final Actions” on page 2-4.