Technical Specifications
Progressa™ Bed User Manual (171528 REV 1) 83
Environmental Conditions for Transport and Storage
Environmental Conditions for Use
Mains Power Requirements
Fuse Specifications
Auxiliary Outlet Power Specifications
Condition Range
Temperature -40°F (-40°C) to 158°F (70°C)
Relative Humidity 10 to 95%
Pressure 500 hPa to 1060 hPa
Condition Range
Ambient Temperature—Progressa™ Prevention
Surface
50° F to 104° F (10° C to 40° C)
Ambient Temperature—Progressa™ Therapy
and Pulmonary Surfaces
50° F to 86° F (10° C to 30° C)
Relative Humidity Range 20% to 85% non-condensing
Atmospheric Pressure 70 kPa to 106 kPa
Altitude Medical electric equipment rated to operate at an
altitude of less than 9842.5' (3000 m)
Condition Range
Rated Voltage 100 V/110 V/115 V/120 V/127 V/220 V/230 V/
240 V AC
Power/Input 6 A (220 V, 230 V, and 240 V beds)
12 A (100 V, 110 V, 120 V, and 127 V beds)
Frequency 60/50 Hz (all beds)
Condition Range
Air System Fuse (air system optional) 2 A, 250 V~, 5 x 20 mm, UL 198G Fast Acting
Battery Fuse 10 A, 32 V~, ATO
Mains Fuse (100V, 110V, 120V, and 127V bed
model)
2 each 15 A, 250 V~, ¼" x 1¼", UL 198G Slo-
Blo® or equivalent
Mains Fuse (220V, 230V, and 240V bed model) 6.3 A, 250 V~, 5 x 20 mm, IEC127 Sheet III,
Time Delay
Condition Range
Receptacle 12 A outlet, electrically isolated from the bed’s
mains power (100 to 137 VAC beds)
6 A outlet, electrically isolated from the bed’s
mains power (220 to 240 VAC beds)