Technical Specifications
86 Progressa™ Bed User Manual (171528 REV 1)
Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The P7500 is intended for use in the electromagnetic environment specified below. The customer or the
user of the model P7500 should make sure it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment—Guidance
RF emissions
CISPR 11
Group 1 The P7500 uses RF energy only for its internal
functions. Therefore, its RF emissions are low and
are not likely to cause any interference in nearby
electronic equipment.
RF Emissions
CISPR 11
Class A The P7500 is suitable for use in all establishments
other than domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The P7500 is intended for use in the electromagnetic environment specified below. The customer or the
user of the P7500 should make sure it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment—
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood,
concrete, or ceramic
tile. If floors are cov-
ered with synthetic
material, the relative
humidity should be at
least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
± 2 kV for Power Sup-
ply Lines
± 1 kV for Input/ Output
Lines
± 2 kV for Power Sup-
ply Lines
± 1 kV for Input/ Output
Lines
Mains power quality
should be that of a typ-
ical commercial or hos-
pital environment.
Surge
IEC 61000-4-5
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to Earth
± 1 kV Line(s) to
Line(s)
± 2 kV Line(s) to Earth
Mains power quality
should be that of a typ-
ical commercial or hos-
pital environment.