Technical Specifications
Progressa™ Bed User Manual (171528 REV 1) 85
Classification and Standards
The Progressa™ Bed is designed and manufactured according to the following equipment
classifications and standards:
Electromagnetic Emissions Guidance
CAUTION:
This device meets all requirements for electromagnetic compatibility per IEC 60601-1-2. It is
unlikely that the user will encounter problems with this device because of inadequate
electromagnetic immunity. However, electromagnetic immunity is always relative, and standards
are based on anticipated environments of use. If the user observes unusual device behavior,
particularly if such behavior is intermittent and associated with nearby use of radio or TV
transmitters, cell phones, or electro-surgical equipment, this could be an indication of
electromagnetic-interference. If such behavior occurs, the user should try to move the interfering
equipment further from this device.
WARNING:
The P7500 should not be used adjacent to or stacked with other electrical equipment. If adjacent
or stacked use is necessary, observe the P7500 and the other electrical equipment to make sure
they operate as intended.
Make sure the P7500 operates correctly when it is used near other electronic devices. Portable
and mobile radio frequency (RF) communications equipment can affect electrical equipment.
Medical equipment needs special precautions in regard to electromagnetic compatibility (EMC)
and needs to be installed and put into service according to the EMC information supplied in the
tables that follow.
Equipment Classification per EN 60601-1
Class I equipment, internally powered equip-
ment
Degree of Protection Against Electric Shock Type B
Classification According to Directive 93/42/EEC
Class I
Class IIa for therapy and pulmonary surfaces
Degree of Protection Against Ingress of Water Ordinary Equipment - IPX4
Degree of Protection Against the Presence of
Flammable Anesthetic Mixtures
Not for use with flammable anaesthetics.
Mode of Operation (Bed Articulation) Continuous operation with intermittent loading,
2 minutes ON/18 minutes OFF
Attenuation (radiolucent panel aluminium equiva-
lencies)
0.179 mm (one panel)
0.191 mm (two panels)