Technical Specifications
Progressa® Bed Instructions for Use (171528 REV 9) 91
Environmental Conditions for Transport and Storage
Environmental Conditions for Use
Mains Power Requirements
Fuse Specifications
Auxiliary Outlet Power Specifications
Applied Parts (in accordance with IEC 60601-1)
Condition Range
Temperature -20°F (-29°C) to 140°F (60°C)
Relative Humidity 15 to 90%
Pressure 500 hPa to 1060 hPa
Condition Range
Ambient Temperature—Progressa® Prevention
Surface
50° F to 104° F (10° C to 40° C)
Ambient Temperature—Progressa® Therapy and
Pulmonary Surfaces
50° F to 86° F (10° C to 30° C)
Relative Humidity Range 20% to 85% non-condensing
Atmospheric Pressure 70 kPa to 106 kPa
Altitude Medical electric equipment rated to operate at an
altitude of less than 9842.5' (3000 m)
Condition Range
Rated Voltage 100 V/110 V/115 V/120 V/127 V/220 V/
230 V/240 V AC
Power/Input 6 A (220 V, 230 V, and 240 V beds)
10 A (100 V, 110 V, 120 V, and 127 V beds)
Frequency 60/50 Hz (all beds)
Condition Range
Air System Fuse (air system optional) 6.3 A, 250 V~, 5 x 20 mm, UL 248-1 Slo-Blo® or
equivalent
Battery Fuse 10 A, 32 V~, ATO
Mains Fuse (100V, 110V, 120V, and 127V bed
model)
2 each 15 A, 250 V~, ¼" x 1¼", UL 248-1 Slo-Blo® or
equivalent
Mains Fuse (220V, 230V, and 240V bed model) 6.3 A, 250 V~, 5 x 20 mm, IEC127 Sheet III,
Time
Delay
Condition Range
Receptacle 12 A outlet, electrically isolated from the bed’s
mains power (120 VAC beds)
Siderail Headboard
Footboard Caregiver Pendant
Sleep deck Sleep surface
Line manager