Safety Information
76 Progressa® Bed Instructions for Use (171528 REV 9)
SAFETY INFORMATION
BED POSITIONS
WARNING:
Warning—Medical bed should be left in its lowest position when the patient is unattended in order to
reduce risk of injury due to falls.
BRAKES
WARNING:
Warning—Always set the brakes when the bed is occupied, except during patient transport. To help
make sure the bed will not move, push and pull on the bed to check it after the brakes are engaged.
Brakes should always be set when the bed is occupied and especially when moving a patient from one
surface to another. Patients often use the bed for support when getting out of bed and could be injured
if the bed unexpectedly moves. After setting the brakes, push and pull the bed to make sure of stability.
Failure to do so could cause injury or equipment damage.
FLUIDS
WARNING:
Warning—Fluid spills onto the bed electronics can cause a hazard. If such a spill occurs, unplug the bed,
and remove it from service. Failure to do so could cause injury or equipment damage.
When fluid spills occur, outside of those seen in normal use, immediately:
Unplug the bed from its power source.
Remove the patient from the bed.
Clean the fluid spill from the bed.
Have maintenance inspect the bed completely.
Do not put the bed back into service until it is completely dry, tested, and determined to be safe to
operate.
SIDERAILS
Siderails may serve several beneficial uses including providing an edge reminder, bed exit assist, and
access to caregiver interface and patient controls. The use of siderails also may provide a sense of
security. Siderails should always be in the upright and latched position when the bed is in the chair
position. The use of siderails in the bed position should be determined according to patient need after
assessing any risk factors according to the facility protocols for safe positioning.
When raising the siderails, a click indicates that the siderails are completely raised and locked in place.
once the click is heard, gently pull on the siderail to make sure the siderail is latched in position.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Progressa P7500 and is the answer not in the manual?

Hillrom Progressa P7500 Specifications

General IconGeneral
BrandHillrom
ModelProgressa P7500
CategoryMedical Equipment
LanguageEnglish

Related product manuals