Surfaces
64 Progressa® Bed Instructions for Use (171528 REV 9)
To Activate
1. Press the Surface button on the GCI.
2. Press Right or Left Turn Assist. The button turns green when active.
To stop Turn Assist, press the Normal control.
To hold the turn at less than the full angle, press the Hold control while the Turn Assist is
inflating.
After 28 minutes, a beep will sound and a screen will appear that there are 2 minutes left. The caregiver
has the option of keeping the surface in Turn Assist or let it return to Normal Mode
If the siderail the patient is turning towards is lowered the turn assist will stop.
To Deactivate
•Press the Normal control.
PROGRESSA® PULMONARY SURFACE
The Progressa® Pulmonary Surface features are the same as the Progressa® Therapy Surface, with the
addition of Rotation, Percussion and Vibration, and Opti-Rest features. Refer to “Progressa® Therapy
Surface” on page 60 for Progressa® Therapy Surface operation.
CONTRAINDICATION:
Contraindication—Use of active air therapy surfaces for patients with unstable spinal cord injury could
cause serious injury to the patient.
Contraindication—Use of continuous lateral rotation therapy is contraindicated for patients with
cervical or skeletal traction.
WARNING:
To help prevent injury and/or equipment damage, obey these warnings:
Warning—Use care when you transfer a patient from the bed to another surface.
WarningOperating Percussion and Vibration and Rotation Therapy together at higher than
typical settings may cause elevated surface temperatures and patient harm, for example, the
combination of the following control settings:
Rotation therapy programmed at 100% with a 1 minute center pause time
Rotation therapy operating continuously
Percussion and Vibration programmed at a high setting
Percussion and Vibration operating for 1 hour periods, greater than the rate of 1 hour for
every 5 hours of Rotation therapy operation
The recommended therapeutic weight range for pressure relief and turning capabilities is 70 to 500 lb
(32 to 227 kg).
The pulmonary surface has a MicroClimate Management® (MCM) topper that operates continuously
while the patient is on the bed and helps decrease localized heat and moisture buildup that occurs
between the patient and the surface.
The surface uses input from the bed scale system to adjust the cushion pressures based on the patient’s
weight.