Graphical Caregiver Interface (GCI)® Controls
30 Progressa® Bed Instructions for Use (171528 REV 9)
HEAD ANGLE ALARM
The head angle alarm lets the caregiver set
an alarm to sound if the head section goes
below 30° or 45°. A message will show on
the GCI when the head section goes below
the angle setting.
To Activate
1. Raise the head section to the
applicable position above 30° or 45°.
2. Press the Alarms control on the GCI.
3. Press the desired head angle alarm.
When an alarm sounds
Raise the head section above 30° or
45°.
or
1. Press the Alarms control on the GCI.
2. Press Off to silence the alarm.
SCALE
The Scale control on the GCI allows you to
Zero the Scale (does not clear history), New
Patient (clears history and zeroes scale),
Weigh Patient, adjust the weight,
add/remove items, change from pounds
(lbs) to kilograms (kg), calculate BMI, or view weight history.
If the bed has a pendant installed, make sure it is either on the siderail or footboard when you zero the
scale or weigh a patient.
Scale specifications
Scale accuracy: 2.2 lb (0.99 kg) or 1% of patient weight, whichever is greater
Scale repeatability: 2.2 lb (0.99 kg) or 1% of patient weight, whichever is greater
The maximum scale capacity is 551 lb (250 kg), however the maximum patient weight for the bed is 500
lb (227 kg).