Information Indicators
14 Progressa® Bed Instructions for Use (171528 REV 9)
A triple beep will sound when there is an error or caregiver attention is needed. A message will appear
on the GCI for further instructions.
BED BATTERY POWER
Charged - The Charged indicator comes on when the battery
is charged.
If the bed is unplugged press any function to activate the
battery power status.
Low - The Low indicator flashes when the battery is low. An
intermittent tone sounds every two minutes when the battery
reaches low condition and AC is unplugged.
Off - If the battery is too low to operate the bed.
Caution—Although a fully-charged battery is preferred, transport may be done when the battery
charge is low. The bed should be reconnected to AC power as soon as possible to prevent equipment
damage.
If the Battery Indicator changes from Charged to Low consistently within four hours of being
disconnected from AC power, the battery should be replaced.
While on battery power, the following will happen:
• All bed articulations will operate
• Integrated surfaces will remain inflated, but will not adjust pressures
• The GCI will not display
SERVICE REQUIRED
The Service required indicator illuminates when the bed detects a malfunction. Contact the
facility maintenance department for assistance.
HIP POSITION LOCATOR
A hip position label appears on the intermediate siderails to indicate
the correct position of the patient's hips while on the bed. The labels
are on the top of the intermediate siderail just above the caregiver
controls.
Proper placement of the patient increases the effectiveness of the
SlideGuard® Patient Position Mechanism and StayInPlace™ Patient
Position Mechanism. These minimize migration of the patient to the
foot end of the bed when you raise the head section.