Technical Specifications
96 Progressa® Bed Instructions for Use (171528 REV 9)
Electromagnetic Immunity Guidance
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The P7500 is intended for use in the electromagnetic environment specified below. The customer or the user of the
P7500 should make sure it is used in such an environment.
Immunity Test
EN and IEC
60601-1-2
4th Edition
Test Level
Compliance
Level
Electromagnetic Environment—Guidance
Conducted
RF-Immunity
IEC 61000-4-6
Radiated
RF-Immunity
IEC 61000-4-3
3 V (80% AM)
150 kHz to
80 MHz
(6V in ISM
Bands per
CISPR-11)
3 V/m
80 MHz to
2700 MHz
3 V (80% AM)
150 kHz to
80 MHz
(6V in ISM
Bands per
CISPR-11)
10 V/m
80 MHz to
2700 MHz
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, should be less than the compli-
ance level in each frequency range.
Interference may occur in the vicinity of equipment marked
with this symbol.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Note 3: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast can not be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the P7500 is used exceeds the applicable
RF compliance level above, the P7500 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the P7500.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Progressa P7500 and is the answer not in the manual?

Hillrom Progressa P7500 Specifications

General IconGeneral
BrandHillrom
ModelProgressa P7500
CategoryMedical Equipment
LanguageEnglish

Related product manuals