Preventive Maintenance
82 Progressa® Bed Instructions for Use (171528 REV 9)
PREVENTIVE MAINTENANCE
WARNING:
To help prevent injury and/or equipment damage, obey these warnings:
Warning—Only facility-authorized personnel should perform calibration on the Progressa®
Bed.
Warning—Only facility-authorized personnel should perform preventive maintenance on the
Progressa® Bed.
The Progressa® Bed requires an effective maintenance program. We recommend that you perform
annual preventive maintenance (PM) and testing for Joint Commission certification. PM and testing not
only meet Joint Commission requirements but can help support a long, operative life for the Progressa®
Bed. PM will minimize downtime due to excessive wear. For the preventive maintenance detailed
procedures, refer to the Progressa® Bed Service Manual (171748).
Perform annual preventive maintenance procedures to make sure all bed components are functioning
as originally designed. Pay particular attention to safety features, including but not limited to:
Siderail latching mechanisms
Siderail dampers for oil leaks
•Caster braking systems
Electrical system components
Electrical power cords for fraying, damage, and proper grounding
All controls return to off or neutral position when released
Controls or cabling entanglement in system mechanisms or siderails
Proper operation of the lockout controls
Integrity of sleep surface cover
Intake/Exhaust filters for cleanliness and condition
Main Battery
Replace the battery if any of these conditions exist (refer to the Progressa® Bed Service Manual (171748):
The battery indicator does not come on within 3 minutes of bed connection to AC mains.
The battery indicator does not increase the number of illuminated LEDs within 12 hours of bed
connection to AC mains.
IntelliDrive® Transport System Batteries
Replace the batteries if the IntelliDrive® Transport System automatically shuts down power before the
final battery charge indication LED flashes (refer to the Progressa® Bed Service Manual (171748)).
Press the blue button on the end of the drive box to disable the battery if the bed will be stored for an
extended period of time.
After replacing the batteries, charge the batteries a minimum of 20 hours before use.
NOTE:
Follow instructions on the batteries for proper disposal or recycling.
Troubleshooting

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Progressa P7500 and is the answer not in the manual?

Hillrom Progressa P7500 Specifications

General IconGeneral
BrandHillrom
ModelProgressa P7500
CategoryMedical Equipment
LanguageEnglish

Related product manuals