Caregiver Siderail Controls
Progressa® Bed Instructions for Use (171528 REV 9) 25
To Activate:
•Press a Nurse Call control.
When the nurse station acknowledges the nurse call, the inboard indicator
continuously illuminates amber and the outboard indicator does not
illuminate.
When the nurse station communication line is open, both the inboard and the
outboard indicators continuously illuminate green.
Speak into the speaker/microphone located on the inboard side of the head
end siderails.
MAX-INFLATE (SIDERAIL METHOD)
To Activate
Press the Max-Inflate control.
The green indicator light will turn on.
To Deactivate
Press the Max-Inflate control
The green indicator light will turn off.
Refer to “Max-Inflate” on page 61 for an alternative method. This feature times out after 30 minutes.
SIDE EXIT ASSIST
The Side Exit Assist control inflates the seat section of the surface to assist in side exit of
the bed. This feature times out after 30 minutes.
To Activate
1. Help patient to side sitting position on edge of mattress.
2. Raise or lower the bed so the patient's feet will be flat on the floor.
3. Press the Side Exit Assist control on the head end siderail.
4. After the seat section inflates, assist the patient with bed exit.
To Deactivate
Press the Side Exit Assist control on the head end siderail.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Progressa P7500 and is the answer not in the manual?

Hillrom Progressa P7500 Specifications

General IconGeneral
BrandHillrom
ModelProgressa P7500
CategoryMedical Equipment
LanguageEnglish

Related product manuals