Technical Specifications
92 Progressa® Bed Instructions for Use (171528 REV 9)
Scale Classification (European Scale Beds Only)
Nurse Call Connection Requirements
For information about the Nurse Call connection requirements, refer to the SideCom®
Communication System Design and Application Manual (DS059).
Mattress Compatibility
Warning—The Envision E700 surface on the Progressa® Bed frame is not fully compliant to the
IEC 60601-2-52:2009 standard; however, it is compliant with the FDA Guidance: Hospital Bed System
Dimensional and Assessment Guidance to Reduced Entrapment [Issued March 10, 2006] standard. Use
of a mattress in combination with the product that is not fully compliant to the IEC 60601-2-52:2009
standard may increase the risk of patients becoming trapped. In such cases, the patient must be
monitored closely.
Warning—The following surfaces can be used with the Progressa® Bed with the Dining Chair® feature
option. Do not use the following surfaces with the Chair Egress option. Do not use the FlexAfoot™
feature with the following surfaces:
• P500 MRS
• NP100—flat deck 36" x 84" (91 cm x 213 cm)
• AccuMax® Surface—flat deck 36" x 84" (91 cm x 213 cm)
• Accella™ Therapy MCM™ P006788A—flat deck 36" x 84" (91 cm x 213 cm)
Condition Range
Technical and Quality Standards EN45501
Classification per EN 45501 Class III
Bed Configuration Available Mattress
Dining
Chair®
Chair
Egress
Chair Egress with
StayInPlace™
Feature
Prevention Therapy Pulmonary
(CLRT only)
Full
Pulmonary
(CLRT and P
&V)
X X X X X
X X X X X
X X X X X