CAUTION Do not use bleach solutions of any kind when cleaning metal electrical
contacts. They will damage the device.
Select a cleaning agent from the following table.
Section 1. Approved for all Connex Spot Monitor components
Cleaning agent Additional information
Accel INTERVention
Accel TB
CaviWipes
Clinell® Universal Wipes
Oxiver TB
Sani-Cloth
®
Plus
Super Sani-Cloth
®
70 percent isopropyl alcohol solution Applied to a clean cloth
Section 2. Not approved for all Connex Spot Monitor components
NOTE The following cleaning agents are NOT approved for cleaning Connex Spot
Monitors equipped with the Braun ThermoScan PRO 6000.
Cleaning agent Additional information
Bacillol
®
AF Wipes
Not approved for use on the display
Cleancide
Clinitex
®
Detergent Wipes
Not approved for use on the display
Clorox Dispatch Wipes
Not approved for use on the display
Clorox Fuzion
Not approved for use on the display
Clorox HealthCare Bleach Germicidal Cleaner
Mikrozid
®
AF Wipes
Not approved for use on the display
Oxivir
®
1 Wipes
Not approved for use on the display
Oxivir Plus 1:40 Solution
Not approved for use on the display
98 Cleaning requirements Welch Allyn® Connex® Spot Monitor