Specify advanced averaging interval settings
Averaging intervals are available in the Office profile.
1. In Advanced settings, touch the Parameters > Averaging tabs.
2. Specify settings.
Setting Action/Description
NIBP averaging program Select the averaging program to view or modify.
You can change the name of the averaging
program.
Readings to average Select which readings in a series to use for
averaging (and the number of readings to average
as a result).
Delay to start (minutes) Enter the number of minutes to wait before the
averaging program starts (after touching "Start
intervals") and the start of the first reading.
Time between (minutes) Enter the number of minutes to wait between
readings.
Keep if + or – (mmHg) Enter the range that the program uses as criteria to
accept or reject readings and establish the baseline
reading. For more information about how this
setting affects the program, refer to the section
"Excluded readings" below.
Summary display box Displays currently selected settings.
3. Do one of the following:
To continue in Advanced settings, touch another tab.
To exit Advanced settings and return to the Home tab, touch Exit.
Specify advanced program interval settings
1. In Advanced settings, touch the Parameters > Program tabs.
2. Specify settings.
Setting Action/Description
Program Select the program to view or modify. You can
change the name of the program.
Intervals Specify up to five sets of interval readings for this
program. For each set, specify the number of
minutes between interval readings and the number
of times that each interval should run (repetitions)
before proceeding to the next set.
3. Do one of the following:
To continue in Advanced settings, touch another tab.
To exit Advanced settings and return to the Home tab, touch Exit.
Network tab
Specify radio settings
This task is applicable only to monitors that have a wireless radio installed. Devices can be
configured with a wireless or a Bluetooth radio.
26 Advanced settings Welch Allyn® Connex® Spot Monitor

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Welch Allyn Connex Spot Monitor and is the answer not in the manual?

Hillrom Welch Allyn Connex Spot Monitor Specifications

General IconGeneral
BrandHillrom
ModelWelch Allyn Connex Spot Monitor
CategoryMedical Equipment
LanguageEnglish

Related product manuals