NOTE After performing any of these procedures and before returning the device
to service, you must use the service tool, Gold edition, to complete the full suite of
functional tests to ensure that all systems are operating within the design
specifications. For more information about these tests and the service tool, see
“Functional verification and calibration.” If you do not have the service tool, you
can download it from https://www.hillrom.com/en/services/welch-allyn-service-
tool/.
Power down the monitor
You can power down the device as follows: 1) by briefly pressing the power button on the housing
and then following the onscreen prompts, or 2) by using onscreen controls alone.
Option 1. Press the power button. An onscreen dialog opens with the following options:
Power down. Device behavior is the same for all profiles.
Sleep. The Sleep button clears the display and puts the device into Display power-saving
mode.
Cancel. The Cancel button dismisses the dialog.
Touch Power down and complete the power-down procedure as described in Option 2, steps 3
and 4.
Option 2. Use onscreen controls alone to power down the device:
1. Touch the Settings tab.
2. Touch the Device tab.
3. Touch Power down.
4. Select the desired option and then touch Power down.
The Power-down options dialog
The buttons in this dialog produce varied effects, as noted below:
Power down. The device clears all data onscreen and performs a complete software
shutdown.
Cancel. This button dismisses the dialog and returns you to the previous screen.
NOTE If your monitor is configured with a Braun thermometer, avoid powering
down the monitor between patient measurements to help maintain an adequate
charge to the Braun dock.
Service manual Disassembly and repair 47

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Welch Allyn Connex Spot Monitor and is the answer not in the manual?

Hillrom Welch Allyn Connex Spot Monitor Specifications

General IconGeneral
BrandHillrom
ModelWelch Allyn Connex Spot Monitor
CategoryMedical Equipment
LanguageEnglish

Related product manuals