The new name appears in the list of paired Bluetooth device list.
Bluetooth Low Energy (BLE) workflow
For more information, contact your sales representative.
Use the Welch Allyn Product Configuration Tool (version 1.9.0, or later) to allow and enable the
Bluetooth Low Energy (BLE) connection and update the Connex Spot Monitor (monitor)
configuration file.
Refer to the “Advanced Settings” instructions on how to allow the Bluetooth Low Energy
configuration.
1. Power on the Connex Spot Monitor.
2. Open the mobile application on the device. A list of Vitals devices appears.
3. Select the Vitals device in the mobile application. If this is the first time connecting the mobile
device with the Connex Spot Monitor or if “Requiring pairing confirmation” is enabled:
a. The Bluetooth Pairing request prompt appears: "WACSM... would like to pair with your ..."
b. Pair the device and the Connex Spot Monitor by touching OK on the Connex Spot
Monitor at the prompt, "A Bluetooth® Low Energy device is attempting to connect."
c. At the pairing confirmation screen, touch Pair on the mobile application.
The mobile application home screen appears.
Service manual Bluetooth wireless technology 39

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Welch Allyn Connex Spot Monitor and is the answer not in the manual?

Hillrom Welch Allyn Connex Spot Monitor Specifications

General IconGeneral
BrandHillrom
ModelWelch Allyn Connex Spot Monitor
CategoryMedical Equipment
LanguageEnglish

Related product manuals