Service and maintenance toolset
Connex Spot Monitor
The following is the listing of tools and test equipment required to perform basic functional
verification checks, full functional verification checks and calibration, and service and maintenance
on the Welch Allyn Connex Spot Monitor. Please refer to the device service manual for detailed
instructions to service your device.
NOTE Tools and test equipment must also be readily available to complete Welch
Allyn Technical Service Training for this device.
Material no. Description Qty
Usage
Full Basic
General test equipment
106270 Service test box 1 X
106275 Micro USB cable 1 X
N/A Wireless router, a/b/g or equivalent 1 X
N/A Ethernet cable 1 X
N/A PC running Windows 10 1 X X
web download Welch Allyn Service Tool 1 X X
1001120 Welch Allyn Service Tool, Gold license 1 X
NIBP testing
407672 BP test volume repair fixture 113670 1 X X
N/A Pressure meter (must include at least two decimal points
and be accurate to within ±0.5 mmHg)
1 X X
N/A Blood pressure Y-tube 1 X X
6000-30 Single tube blood pressure hose, 5 feet 1 X X
253

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Welch Allyn Connex Spot Monitor and is the answer not in the manual?

Hillrom Welch Allyn Connex Spot Monitor Specifications

General IconGeneral
BrandHillrom
ModelWelch Allyn Connex Spot Monitor
CategoryMedical Equipment
LanguageEnglish

Related product manuals