Material no. Description Qty
Usage
Full Basic
4500-30 Blood pressure hose, 5 ft. 1 X
620216 Fitting "Y" 1/8 X 1/8 X 1/8 1 X X
SpO2 configuration testing
N/A SpO2 simulator 1 X X
DOC-10 Nellcor SpO2 extension cable 1 X X
SureTemp Plus configuration testing
06138-000 Cal-key, assembly, M690/692 1 X X
01802-110 Tester, calibration, 9600 Plus 3 X
Braun ThermoScan PRO 6000 configuration testing
01802-110 Tester, calibration, 9600 Plus 3 X
Notes:
All items marked "N/A" in the Material no. column are OEM or off-the-shelf tools and equipment not
available from Welch Allyn. The remaining items are available through Welch Allyn or a Welch Allyn
distributor.
All items marked in the Full column are required for full functional verification checks and calibration.
The Welch Allyn Service Tool software and an active Gold license is also required to perform these
checks. The Gold license is obtained upon successful completion of Welch Allyn Technical Service
Training.
All items marked in the Basic column are required for basic functional verification checks. The Welch
Allyn Service Tool software Standard unlicensed edition can be used to perform the NIBP checks and
SWF upgrades/updates.
Miscellaneous Tools
Phillips #1 screwdriver
T-10 Torx screwdriver
3mm hex key
254 Service and maintenance toolset Welch Allyn® Connex® Spot Monitor

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Welch Allyn Connex Spot Monitor and is the answer not in the manual?

Hillrom Welch Allyn Connex Spot Monitor Specifications

General IconGeneral
BrandHillrom
ModelWelch Allyn Connex Spot Monitor
CategoryMedical Equipment
LanguageEnglish

Related product manuals