Cleaning agent Additional information
Reynard Neutral Detergent Wipes Not approved for use on the display
Reynard Premier Disinfectant Wipes Not approved for use on the display
Sani-Cloth Active Wipes Not approved for use on the display
Sani-Cloth
®
Bleach
Not approved for use on the display
Sani-Cloth
®
Prime Wipes
Not approved for use on the display
Sekusept™ Plus 1.5% Solution Not approved for use on the display
Super HDQ
®
L10
Dilution rate of ½ oz per gallon of water (1:256) applied
to a clean cloth
Tuffie5 Cleaning Wipes
Viraguard Wipes Not approved for use on the display
Virex II (256) Dilution rate of ½ oz per gallon of water (1:256) applied
to a clean cloth
10 percent bleach solution (.5% - 1% sodium hypochlorite) applied to a clean cloth
Remove liquid spills from the monitor
Liquids can damage electronics inside the monitor. Follow these steps if liquids spill on the
monitor.
1. Power down the monitor.
2. Disconnect the power cord from the mains outlet and the power source.
3. Remove battery pack from the monitor.
4. Dry excess liquid from the monitor.
5. Reinstall battery pack.
6. Reconnect the power cord.
7. Power on the monitor and verify that the monitor functions normally before using it.
If liquids possibly entered the monitor, remove the monitor from use until it has been properly
dried, inspected, and tested by qualified service personnel.
Clean the equipment
The screen lock blocks the display of patient information and prevents any input, which may be
useful when cleaning the display.
Follow the cleaning agent manufacturer's instructions to prepare solution, if applicable, and clean
all exposed surfaces of the monitor, APM work surface, accessory bin(s) and basket, cords and
cables, and stand. Wipe all surfaces until no visible soil remains. Change the wipe or cloth
throughout the cleaning procedure as needed.
Service manual Cleaning requirements 99

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Welch Allyn Connex Spot Monitor and is the answer not in the manual?

Hillrom Welch Allyn Connex Spot Monitor Specifications

General IconGeneral
BrandHillrom
ModelWelch Allyn Connex Spot Monitor
CategoryMedical Equipment
LanguageEnglish

Related product manuals