1. In Advanced settings, touch the Network > Radio tabs.
2. Specify settings.
Setting Action/Description
Enable radio Enable the radio for WiFi communications. When
disabled, the radio is not available.
ESSID Enter the service set identifier (SSID) for the
enterprise. Enter a maximum of 16 characters.
Radio mode Select the radio band.
Dynamic Frequency Selection Enable to allow the site to transmit data on
changing frequencies for added security. This
option applies most frequently to military or
government enterprises.
Enable radio network alarms Activate radio network alarms when an wireless
dropout alarm condition occurs. When disabled,
radio network alarms are not available.
Roaming Select the type of roaming (only available with
WPA/WPA2 Enterprise).
Regulatory Display the WiFi regulatory certification
information.
Update radio Touch Update radio to activate all new radio
settings not selected previously.
Touch OK in the confirmation popup.
NOTE None of the changed
radio settings take effect until
you touch Update radio.
3. Do one of the following:
To continue in Advanced settings, touch another tab.
To exit Advanced settings and return to the Home tab, touch Exit.
Specify security settings
1. In Advanced settings, touch the Network > Security tabs.
2. Select the encryption method to secure data transfer from the device.
NOTE Network server certificates are required for all EAP security
options. Use the Welch Allyn Service Tool to load these certificates.
3. Specify settings. Do one of the following:
Setting Action/Description
Authentication type Select the preferred encryption option. Then
specify any additional settings that appear. The
default is Open or no encryption.
WEP 64 Select a WEP key and then enter a 10-character key
in the selected field. Repeat this process to create
multiple WEP keys. Then click Update radio.
WEP 128 Select a WEP key and then enter a 10-character key
in the selected field. Repeat this process to create
multiple WEP keys. Then click Update radio.
Service manual Advanced settings 27

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hillrom Welch Allyn Connex Spot Monitor and is the answer not in the manual?

Hillrom Welch Allyn Connex Spot Monitor Specifications

General IconGeneral
BrandHillrom
ModelWelch Allyn Connex Spot Monitor
CategoryMedical Equipment
LanguageEnglish

Related product manuals