you are wearing a grounded wrist strap (with a resistor of at least 1 megohm in series) or other
grounding device.
Use only grounded tools when inserting, adjusting, or removing static-sensitive components
and assemblies.
Remove or insert static-sensitive components and assemblies only with monitor power turned
off.
Insert and seal static-sensitive components and assemblies into their original static-shielding
bags before removing them from static-protected areas.
Always test your ground strap, bench mat, conductive work surface, and ground cord before
removing components and assemblies from their protective bags and before beginning any
disassembly or assembly procedures.
Service manual Safety 7