93
────────────────────────────────────────────────────
7.4 Command
────────────────────────────────────────────────────
1
2
3
4
5
7
9
10
11
12
13
14
A
Sets the comparative voltage value.
Syntax
:CONF:VOLT <data>
data
0.00 - 5.00
( Numerical data in NR2 format, three digits )
The comparative voltage value ( Refer to section 7.5. )
Response
OK
Function
In the READY or SETTING state, set the comparative-voltage value.
Example
Sets the comparative voltage value to 1.50 kV.
PC
>
3158
>
:CONF:VOLT 1.50
OK Setting completion.
Error
The execution of this command in a state other than the READY and SETTING state
causes an execution error.
Queries the comparative voltage value enablement.
Syntax
:CONF:VOLT?
Response
<data>
data
Comparative voltage value ( Refer to section 7.5. )
Function
Queries the comparative voltage value enablement.
Example
PC
>
3158
>
:CONF:VOLT?
1.50 The comparative voltage value is 1.50 kV.
:CONF:VOLT
:CONF:VOLT?